Clarity of solution
Hydrazine sulfate solution
Transfer 1.0 g of hydrazine sulfate to a 100-mL volumetric flask, dissolve in and dilute with water to volume, and mix. Allow to stand for 4 to 6 hours before use.
Methenamine solution
Transfer 2.5 g of methenamine to a 100-mL glass-stoppered flask, add 25 mL of water, insert the glass stopper, and mix to dissolve.
Primary opalescent suspension
[noteThis suspension is stable for 2 months, provided it is stored in a glass container free from surface defects. The suspension must not adhere to the glass and must be well mixed before use.] Transfer 25.0 mL of Hydrazine sulfate solution to the Methenamine solution in the 100-mL glass-stoppered flask. Mix, and allow to stand for 24 hours.
Opalescence standard
[noteThis suspension should not be used beyond 24 hours after preparation.] Transfer 15.0 mL of the Primary opalescent suspension to a 1000-mL volumetric flask, dilute with water to volume, and mix.
Reference suspensions
Transfer 5.0 mL of the Opalescence standard to a 100-mL volumetric flask, dilute with water to volume, and mix to obtain Reference suspension A. Transfer 10.0 mL of the Opalescence standard to a second 100-mL volumetric flask, dilute with water to volume, and mix to obtain Reference suspension B.
Test solution
Dissolve 50 mg of Pancuronium Bromide in about 20 mL of water, dilute with water to 25 mL, and mix.
Procedure
Transfer a sufficient portion of the
Test solution to a test tube of colorless, transparent, neutral glass with a flat base and an internal diameter of 15 to 25 mm to obtain a depth of 40 mm. Similarly transfer portions of
Reference suspension A, Reference suspension B, and water to separate matching test tubes. Compare the
Test solution, Reference suspension A, Reference suspension B, and water in diffused daylight, viewing vertically against a black background (see
Visual Comparison under
Spectrophotometry and Light-Scattering 851).
[noteThe diffusion of light must be such that
Reference suspension A can readily be distinguished from water, and
Reference suspension B can readily be distinguished from
Reference suspension A.
] The
Test solution shows the same or more clarity than
Reference suspension A.
Color of solution
Standard stock solution
Prepare a solution of ferric chloride CS, cobaltous chloride CS, cupric sulfate CS, and a 10 g per L hydrochloric acid solution (3:3:2.4:1.6).
Standard solution
[notePrepare this solution just before use.] Transfer 1 mL of Standard stock solution to a 100-mL volumetric flask, dilute with hydrochloric acid (10 g per L) solution to volume, and mix.
Test solution
Use the Test solution prepared as directed in the test for Clarity of solution.
Procedure
Transfer a sufficient portion of the
Test solution to a test tube of colorless, transparent, neutral glass with a flat base and an external diameter of 15 to 25 mm to obtain a depth of 40 mm. Similarly transfer a portion of the
Standard solution and water to separate matching test tubes. Compare the
Test solution (see
Visual Comparison under
Spectrophotometry and Light-Scattering 851): the
Test solution is not more intensely colored than the
Standard solution or water.
Related compounds
Adsorbent:
0.2-mm layer of chromatographic silica gel mixture.
Test solution
Prepare a solution of Pancuronium Bromide in methylene chloride containing 10 mg per mL.
Diluted test solution
Dilute 1.0 mL of the Test solution with methylene chloride to 50 mL, and mix. Dilute 1.0 mL of the resulting solution with methylene chloride to 20 mL, and mix.
Developing solvent system:
a mixture of isopropyl alcohol, acetonitrile, and a 400 g per L solution of sodium iodide (85:10:5).
Procedure
Apply separately 5-µL portions of the
Test solution, Diluted test solution, Standard solution 1, and
Standard solution 2 to a suitable thin-layer chromatographic plate (see
Thin-Layer Chromatography under
Chromatography 621). Develop in an unlined and unsaturated tank over a path of 8 cm. Spray the plate with a 20 g per L solution of sodium nitrite, and allow to dry for 5 minutes. Spray the plate with Dragendorff's TS, and cover the plate with a transparent glass cover. Any spot in the chromatogram obtained from the
Test solution due to vecuronium bromide is not more intense than the corresponding spot in the chromatogram obtained from
Standard solution 1: equivalent to not more than 1.0% of vecuronium bromide. Any other spot in the chromatogram obtained from the
Test solution, except for the principal spot and any spot due to vecuronium bromide, is not more intense than the spot in the chromatogram obtained from the
Diluted test solution: equivalent to not more than 0.1% of any individual impurity. In a valid test, the chromatogram obtained from
Standard solution 1 shows two clearly separated spots. The
RF values for pancuronium bromide and vecuronium bromide are about 0.5 and 0.64, respectively.
Assay
Transfer about 200 mg of Pancuronium Bromide, accurately weighed, to a 100-mL beaker, and dissolve by stirring in 50 mL of acetic anhydride. Titrate with 0.1 N perchloric acid VS, determining the endpoint potentiometrically. Perform a blank determination, and make any necessary correction (see
Titrimetry 541). Each mL of 0.1 N perchloric acid is equivalent to 36.63 mg of C
35H
60Br
2N
2O
4.